In accordance with GMP requirements, it is necessary to check the integrity of the membrane-based filter elements in order to ensure the quality of manufactured products.
Technofilter RME has developed the first Russian automatic instrument for integrity control. This device is extremely easy to operate and allows you to control integritiy of membrane-based filter elements according to the main standard methods.
TechnoCheck®-2 features and benefits
- Accuracy, fast and reliable measurement.
- Operational control and indication of results on touch control panel.
- Fast program selection and data input.
- Easy data reading due to clear display (5,7").
- User – friendly menu design and online hints facilitate the user operation.
- The integrated printer for printing test results ensures data storage.
- Reporting of measurement results in automatic mode using compact printer.
- The change function enables changing programmed test parameters.
- Freely programmable test allows testing filters produced by different manufacturers according to individual integrity test parameters.
- High reliability of the device ensures long service life.
- Test results storage on an external USB-flesh drive.
Measured parameters
- Bubble Point Test
- Diffusive Flow Test
- Pressure Drop Test
Technical specifications
Physical parameters |
|
---|---|
Housing material |
Plastic |
Weight |
7 kg |
Dimensions |
430 x 310 x 170 mm |
mmTouch screen size |
5,7" TFT |
Power requirements |
|
Power supply |
220 V, 50 Hz |
Input power |
300 W |
Protection rating |
IP 54 |
Measuring ranges |
|
Pressure drop |
0-1000 kPa |
Diffusion |
0-300 ml / min |
Bubble point |
0,80-5,50 bar |
Measuring Accuracy |
|
Volume |
±5% |
Pressure |
±0,25% |
Diffusion |
±5 % |
Bubble point |
±0,05 bar |
Operating Conditions |
|
Maximum air pressure |
6 bar |
Ambient temperature |
15-35°С |
Relative humidity |
10-80% |
Storage temperature |
0-40°С |
Certification
Each device is provided with calibration and functionality certificate and also copy of national reference sample (master pressure gauge) calibration certificate. Technofilter provides services on annual calibration of the device following by the issue of an appropriate certificate.
Supply package
- Power cable
- Pressure reducing valve and compressed air filter
- Paper for the integrated dot printer
- Operating instructions, technical documents
Optional equipment
For testing procedure, on agreement with customer, it is possible to equip the device with an optional DS filter housing (or a cell for flat disk membranes) fitted with a tube and a quick-release coupling, reference vessel and a receiver for operation with an individual compressor.